Medical Physics · 1999 · 44 citations · 9 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage AnalysisImage QualityImage Quality AssessmentMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingImaging Result Quality AssuranceMagnetic Resonance Imaging AcquisitionMedical Image Quality AssessmentQuality AssessmentSliceSlice Thickness ControlSpatial Resolution MeasurementUniformity Testing
The authors have developed a protocol and software for the quality assessment of MRI equipment with a commercial test object. Automatic image analysis consists of detecting surfaces and objects, defining regions of interest, acquiring reference point coordinates and establishing gray level profiles. Signal-to-noise ratio, image uniformity, geometrical distortion, slice thickness, slice profile, and spatial resolution are checked. The results are periodically analyzed to evaluate possible drifts with time. The measurements are performed weekly on three MRI scanners made by the Siemens Company (VISION 1.5T, EXPERT 1.0T, and OPEN 0.2T). The results obtained for the three scanners over approximately 3.5 years are presented, analyzed, and compared.
9
Measuring signal-to-noise ratios in MR imaging.
Leo Kaufman, David Kramer, L E Crooks et al. · Radiology · 1989 · 362 citations
Correction of Spatial Distortion in MR Imaging