Concepedia

Publication | Open Access

Comparison of the threshold for peripheral nerve stimulation during gradient switching in whole body MR systems

49

Citations

13

References

2002

Year

Abstract

Then expressed in appropriate format, the thresholds for peripheral nerve stimulation in volunteers for whole body MR equipment can be described with a hyperbolic threshold curve with rheobase 18.8 +/- 0.6 Tesla/second and chronaxie 0.36 +/- 0.02 milliseconds.

References

YearCitations