An Improved Method of Data Collection for 4πβ–γ Coincidence Measurements

David L. Smith

Metrologia · 1975 · 70 citations · 5 references

Concepts

Abstract

A method of data collection for the determination of absolute disintegration rates by the 4 πβ - γ coincidence technique is described, which gives nearly an order of magnitude smaller variance, for a given total counting time, than conventional methods. It is applicable when the β counter output is proportional to the β particle energy. The principle, which is demonstrated for 134Cs, is to record the amplitudes of all the β pulses, followed by integration over variable limits. Expressions for dead time and resolving time corrections are derived.