Metrologia · 1975 · 70 citations · 5 references
EngineeringPhysicsInstrumentationNuclear InstrumentationNuclear PhysicsBeta Particle DetectionCoincidence CountingDead Time CorrectionNuclear DecayParticle DetectorsRadioactive DecayBeta-gamma CoincidenceCoincidence Dead Time CorrectionCoincidence MeasurementsDead Time EffectsVariable Limits Of Integration
A method of data collection for the determination of absolute disintegration rates by the 4 πβ - γ coincidence technique is described, which gives nearly an order of magnitude smaller variance, for a given total counting time, than conventional methods. It is applicable when the β counter output is proportional to the β particle energy. The principle, which is demonstrated for 134Cs, is to record the amplitudes of all the β pulses, followed by integration over variable limits. Expressions for dead time and resolving time corrections are derived.