Publication | Closed Access
Development of a computed tomographic scanner for radiation therapy treatment planning.
11
Citations
0
References
1980
Year
Computed TomographyLow-cost Ct ScannerRadiation MedicineNuclear MedicineMedical ImagingTreatment PlanningComputed Tomographic ScannerRadiotherapy SimulatorCt ScanRadiation Therapy PlanningRadiographic ImagingInstrumentationMedicineRadiation OncologyTreatment VerificationRadiologyHealth Sciences
The authors describe a low-cost CT scanner integrated with a radiotherapy simulator and designed for treatment planning. The standard rotational gantry and x-ray tube of the simulator are used with a multiwire xenon lonization chamber and simple current-proportional readout system to measure patient attenuation, avoiding problems associated with diagnostic CT scanners in treatment planning. Although design constraints limit performance, software compensation techniques have reduced artifacts and given satisfactory images.