British Journal of Radiology · 1990 · 13 citations · 7 references
MedicinePhysicsMedical PhysicsRadiation DosimetryRadiation MeasurementElectron Beam PhysicsElectron DosimetryRadiation Dose MeasurementScatter CorrectionSmall Field DosimetryTreatment Planning Dose CalculationClinical Electron DosimetryElectron Dose CalculationElectron Dose DistributionRadiation Field DesignSmall Field Electron Dosimetry
A simple approach to the calculation of dose in cone-collimated electron fields is presented. The method accounts for variations in lateral scatter with field size using a Clarkson Integration. The reduction in output by an irregularly shaped aperture is evaluated based on measurements of the output at various distances from the aperture in a field 2.5 cm in diameter. The first 14 months of clinical use were analysed and it is shown that in 42 out of 600 calculations, the cutout factor was less than 0.95 and that in 77% of these, the calculation method was accurate to +/- 3%.