Concepedia

Publication | Closed Access

Comparison of cage application modality in posterior lumbar interbody fusion with posterior instrumentation—A finite element study

70

Citations

20

References

2008

Year

Abstract

Posterior instrumentation decreases at least half the distortion stress of cage-endplate interface and facet joints, and diminishes the differences between the stresses of inserting one or two cages. Based on one cage insertion, adding posterior instrumentation provides more efficient stability than an additional cage. Furthermore, an obliquely inserted cage with posterior instrumentation produced lower stress than a cage inserted on one side due to better structural symmetry. Conclusively, one oblique anterior cage and bilateral posterior pedicle screws reconstructed the tripod system as the intact disc and facet joints, and provided similar stability as two cages did.

References

YearCitations