Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs

31

Citations

22

References

2013

Year

Abstract

Excellent performances in terms of dose homogeneity were obtained using a new blue channel method for marker-dye correction on both EBT2 and EBT3 GafchromicTM films. In comparison with TLD, the passing rates for the EBT2 film were higher than for EBT3; a good agreement with estimated data by Monte Carlo algorithm was found for both films, with some statistically significant differences again in favor of EBT2. These results suggest that the use of GafchromicTM EBT2 and EBT3 films is appropriate for dose verification measurements in VMAT stereotactic radiosurgery; taking into account the uncertainty associated with Gafchromic film dosimetry, the use of adequate action levels is strongly advised, in particular, for EBT3.