The Analyst · 1986 · 22 citations · 0 references
Life SciencesBiochemistryEnzymologyEnzymology Of FoodsGlycobiologyBioanalytical ChemistryBioanalytical Method ValidationEnzyme AssaysEnzyme CharacterizationEnzyme DiagnosticsAutomated Enzyme AnalyzersEnzymatic Carbohydrate AssaysEnzymatic Sugar AssaysEnzyme ActivityEnzyme Assay DevelopmentFlow AnalysisSpectrophotometric Enzyme AssaysSpectrophotometric Sugar Assays
An automated flow injection analysis (FIA) method has been developed, using as a chromophore p-nitrophenyl-α-D-glucopyranoside, which is hydrolysed to glucose and p-nitrophenol by amyloglucosidase. The p-nitrophenol is determined spectrophotometrically at 400 nm as a measure of the enzyme activity. The general conditions of the analysis have been optimised. The analytical method is flexibile, as all methods for the analysis of enzymes using soluble p-nitrophenyl derivatives as chromophores may in principle be carried out with the FIA method described. The final result is a simple, automated method with high precision, sensitivity and speed.