Journal of Microscopy · 1998 · 97 citations · 13 references
Natural SciencesBiomedical EngineeringBiomedical ImagingBiomedical OpticsOpticsCellular ImagingConfocal MicroscopyFluorescence ImagingFluorescence MicroscopyLight MicroscopyOptical ImagingConfocal Fluorescence MicroscopyPseudorandom GeneratorsSpatial Light Modulators
The programmable array microscope (PAM) uses a spatial light modulator (SLM) to generate an arbitrary pattern of conjugate illumination and detection elements. The SLM dissects the fluorescent light imaged by the objective into a focal conjugate image, I c , formed by the ‘in‐focus’ light, and a nonconjugate image, I nc , formed by the ‘out‐of‐focus’ light. We discuss two different schemes for confocal imaging using the PAM. In the first, a grid of points is shifted to scan the complete image. The second, faster approach, uses a short tiled pseudorandom sequence of two‐dimensional patterns. In the first case, I c is analogous to a confocal image and I nc to a conventional image minus I c . In the second case I c and I nc are the sum and the difference, respectively, of a conventional and a confocal image. The pseudorandom sequence approach requires post‐processing to retrieve the confocal part, but generates significantly higher signal levels for an equivalent integration time.
13
M. Golay · IEEE Transactions on Information Theory · 1961 · 1.5K citations
New Reflected-Light Microscope for Viewing Unstained Brain and Ganglion Cells
Marcel Egger, Mojmír Petráň · Science · 1967 · 235 citations
Medicine, Neurobiology, Neuroscience +17
Efficient real-time confocal microscopy with white light sources
R. Juŝkaitis, Timothy D. Wilson, Mark A. A. Neil et al. · Nature · 1996 · 140 citations