Applied Physics Letters · 1973 · 12 citations · 1 references
EngineeringEducational MeasurementInstrumentationMetrologySensorsAnalytical CalibrationCalibrationEquipment CalibrationPressure MeasurementTemperature MeasurementTemperature MetrologyAbsolute PressureAbsolute Pressure MeasurementAbsolute Pressure SensorsAtmospheric PressureAtmospheric Pressure MeasurementHigh-pressure MeasurementPressure CalibrationSecondary Pressure Standards
Based on the simultaneous measurement of length and transit time of a specimen under pressure, we have developed a new way of determining absolute high pressure. The mercury freezing point at 0 °C has been determined with this new method at 7571.2±1.6 bar. Accurate high-pressure calibration of a secondary-pressure gauge such as a manganin gauge is possible with this method. The maximum deviation from linearity of the present gauge studied is 11.6±0.1 bar between atmospheric pressure and the mercury freezing point at 0 °C.
1
Measurements in the High-Pressure Environment
Edward C. Lloyd, C.W. Beckett, F. R. Boyd · Science · 1969 · 13 citations