Blood Glucose Measurements by Photoacoustics
28
Citations
10
References
1998
Year
Unknown Venue
Blood Glucose MeasurementsEngineeringPhysiological RangeMedicineIn Vivo SystemDiabetesPhysiologyBiomedical PhotonicsBioanalysisPhotoacoustic ImagingBlood Glucose MonitoringBiophotonicsBiomedical EngineeringClinical ChemistryGlucose ConcentrationsMedical InstrumentationBiophysicsPulsed Photoacoustic Technique
Glucose concentrations have been measured by the pulsed photoacoustic technique both in-vitro and in-vivo. The in- vitro results show that glucose concentrations in the physiological range (30 - 600mg/dl) can be measured with an accuracy of better than 0.1% with a correlation of R 2 = 0.989. The role of other blood analytes in photoacoustic measurements have been explored, ¡η-vivo measurements on 8 subjects have been achieved with a standard deviation of between ± 1.5% and ± 12 % with an R 2 = 0.959 over the glucose concentration range 80 - 640mg/dl.