Size‐specific, scanner‐independent organ dose estimates in contiguous axial and helical head CT examinations
23
Citations
25
References
2014
Year
For all scan modes, strong correlation exists between CTDIvol normalized brain dose and both geometric and attenuation-based patient size metrics. A slightly lower correlation between CTDIvol normalized organ dose and patient size was observed for the lens of the eye. This may be due to the combination of the eye lens being a small peripheral organ and the presence of surface dose variation in both contiguous axial and helical scanning. Results indicate that robust estimates of patient-specific head CT dose may be provided using the size-specific, scanner-independent CTDIvol-to-organ-dose conversion coefficients described in this work.