A Novel Interface for Hybrid Mock Circulations to Evaluate Ventricular Assist Devices
97
Citations
21
References
2012
Year
Heart FailureEngineeringDevice TherapyNovel Mock CirculationWearable TechnologyBiomedical EngineeringFlow ProbeArtificial OrganHybrid Mock CirculationsBlood Flow MeasurementCardiologyCardiac MechanicAssisted CirculationAssistive TechnologyBiomedical FlowNovel InterfaceMechanical Circulatory SupportElectrophysiologyHuman Blood CirculationMedicineVentricular Assist DevicesAnesthesiology
This paper presents a novel mock circulation for the evaluation of ventricular assist devices (VADs), which is based on a hardware-in-the-loop concept. A numerical model of the human blood circulation runs in real time and computes instantaneous pressure, volume, and flow rate values. The VAD to be tested is connected to a numerical-hydraulic interface, which allows the interaction between the VAD and the numerical model of the circulation. The numerical-hydraulic interface consists of two pressure-controlled reservoirs, which apply the computed pressure values from the model to the VAD, and a flow probe to feed the resulting VAD flow rate back to the model. Experimental results are provided to show the proper interaction between a numerical model of the circulation and a mixed-flow blood pump.