Use of recombinant human cardiac Troponin T for standardization of third generation Troponin T methods
100
Citations
2
References
1999
Year
Cardiac MuscleTranslational MedicineHeart FailureLinearity ProblemsTroponin TCardiovascular DiseaseLinear Calibration CurveBioanalysisPhysiologyBiostatisticsPublic HealthMedicineCardiologyCardiac Pathology
The Elecsys Troponin T third generation assay uses recombinant human cardiac troponin T as standard material. The new assay has a linear calibration curve. Thus, linearity problems observed with the second generation assay have been eliminated. The assay has high precision, especially at the low end of measuring range (inter-assay CV < 10% at 0.1 microgram/L). The new standardization does not change the cut-off value of 0.1 microgram/L. The use of recombinant human cardiac troponin T as standard material enables a reproducible and reliable standardization of troponin T assays.