Comparison of the VNIIFTRI and LPTF frequency multiplication chains has been carried out through the measurement of the frequency of a portable VNIIFTRI (He-Ne)/CH4 laser. Agreement is within 100 Hz (1.1 parts in 1012) and is secured by the very good medium-term frequency repeatability of the (He-Ne)/CH4 VNIIFTRI portable laser (a few parts in 1013). On the same occasion a measurement of the frequency of the BIPM (He-Ne)/CH4 reference laser (B.3) has been performed at LPTF. Other experiments carried out on the BIPM laser show that the reproducibility of the (He-Ne)/CH4 system could be improved by a systematic study and then by a better control of the various perturbing factors which influence the shape of the methane-saturated absorption peak.