British Journal of Radiology · 1953 · 27 citations · 1 references
EngineeringNuclear MedicineRadiation DetectionRadiation DosimetryRadiation MeasurementDosimetry InstrumentationGamma-ray DetectionPhoton DosimetryRadiation Dose MeasurementScintillation DetectorsScintillator MaterialsMedical Dosimetry ApplicationsScintillationScintillation DetectionScintillation DosimetryScintillation MaterialsScintillation SpectrometersScintillation SpectroscopyScintillator Crystals
The theoretical considerations influencing the design of scintillation dosemeters are discussed. Details are given of simple and robust probe-type scintillation dosemeters suitable for clinical application. These instruments permit the measurement of dosage-rates due to X or γ radiation of 0·5 per hour and upwards, with a probe having an outside diameter no greater than ¼ in. Experimental data are presented concerning the sensitivity, stability, energy-dependence, temperature-dependence and directional properties of dosemeters of this type incorporating various crystalline luminophors.
1
<i>Some Aspects of the Luminescence of Solids</i>
F. A. Kröger, F. Urbach · Physics Today · 1948 · 418 citations