Publication | Open Access
Changes in the Position of Epidural Catheters Associated with Patient Movement
140
Citations
10
References
1997
Year
Epidural catheters moved a clinically significant amount with reference to the skin in all BMI groups as patients changed position. If catheters had been secured to the skin before position change, many would have been pulled partially out of the epidural space. To minimize the risk of catheter displacement, particularly in obese patients, we recommend that multiorificed catheters be inserted at least 4 cm into the epidural space and that patients assume the sitting upright or lateral position before securing the catheter to the skin.