K- and Ka-band waveguide microcalorimeters for microwave power standards

Tae Ween Kang, Nak Sam Chung, R. Honigbaum, Jürgen Ruhaak, U. Stumper

IEEE Transactions on Instrumentation and Measurement · 1997 · 15 citations · 3 references

Concepts

Abstract

Three sets of waveguide microcalorimeters of K- and Ka-band are described. Included are construction details of the microcalorimeters, operation procedure, and uncertainty analysis. The results of measurements and intercomparisons are presented. The uncertainty analysis shows a total uncertainty including systematic uncertainty and random uncertainty of 0.53-0.68%.

References

3