When a hole slightly smaller than the implant diameter is prepared for implant placement, force-fitting stress increases installation torque and stability can be induced. Thus, large surface roughness of implants should not be viewed as an exclusive mechanism for providing a desirable level of initial fixity. Smaller roughness with the same mean diameter is equally effective.