Artificial Organs · 1996 · 20 citations · 3 references
MedicineCardiologySurgeryCardiothoracic SurgeryHemodynamic ManagementVascular SurgeryCardiopulmonary BypassExtracorporeal CirculationHemodynamic SupportMechanical Circulatory SupportThrombosis PreventionAntithrombin FunctionCoagulationPumpsThrombin Time
Centrifugal pumps have been used widely as the main pump in open heart surgery to reduce damage to blood elements and to reduce the activation of the coagulation system. The purpose of this study was the evaluation and comparison of the effects of two types of centrifugal pumps and of one type of roller pump on blood elements, the coagulation system, complements, and immunoglobulins. Two types of centrifugal pumps (Lifestream; St. Jude Medical, Chelmsford, Massachusetts; and BP-80: Medtronic, BioMedicus, Inc., Eden Prairie, Minnesota, U.S.A.) and one roller pump (Mera Co.) were used separately as the main pump for cardiopulmonary bypass (CPB) in 29 patients. Platelet counts, lactate dehydrogenase, antithrombin III, thrombin-antithrombin complex (TAT), complements (C3, C4, and CH50) and immunoglobulins G, A, and M values were measured before and after CPB and compared. Values, except those for TAT, showed no significant difference among the three groups. The TAT values increased less in each of the centrifugal pump groups than in the roller pump group. This finding suggests that thrombin synthesis might be suppressed by the use of a centrifugal pump.
3
Hemolysis. A comparative study of four nonpulsatile pumps.
T. Oku, H. Harasaki, Wendy D. Smith et al. · PubMed · 1989 · 50 citations
Cardiopulmonary Bypass Using Centrifugal Pump
Turan Berki, Ali Gürbüz, Hakan Akkaya et al. · Vascular Surgery · 1992 · 19 citations
Medicine, Cardiology, Surgery +13
Reoperation and the centrifugal pump?
Heinz Jakob, G. Hafner, Stein Iversen et al. · European Journal of Cardio-Thoracic Surgery · 1992 · 11 citations