Journal of Refractive Surgery · 2012 · 14 citations · 18 references
MedicineOphthalmologySurgeryOphthalmic SurgeryOpticsRefractive SurgeryBiomedical Laser ApplicationsCorneal ImagingCorneal OpticsLaser SurgeryClinical Beam CalibrationCorneal Ablation DepthCorneal Ablation ProfileCorneal Surface AblationExcimer LasersLasersRadiant ExposureSurface Temperature
PURPOSE: To investigate the relevance of initial temperature of the polymethylmethacrylate (PMMA) plates used as a target for photoablation during calibration of excimer lasers performed in daily clinical routine. METHODS: An experimental argon fluoride excimer laser with a repetition rate of 1050 Hz, a radiant exposure of 500 mJ/cm², and single pulse energy of 2.1 mJ was used for photoablation of PMMA plates. The initial plate temperature varied from 10.1°C to 75.7°C. The initial temperature was measured with an infrared camera and the central ablation depth of a myopic ablation of -9.00 diopters (D) with an optical zone of 6.5 mm was measured by means of a surface profiling system. RESULTS: The ablation depth increased linearly from 73.9 to 96.3 μm within a temperature increase from 10.1°C to 75.7°C (increase rate of 0.3192 μm/K). The linear correlation was found to be significant (P<.05) with a coefficient of determination of R²=0.95. Based on these results and assuming a standard room temperature of 20°C, optimal plate temperature was calculated to be 15°C to 25°C to maintain an ablation within 0.25 D. CONCLUSIONS: The temperature of PMMA plates for clinical laser calibration should be controlled ideally within a range of approximately ±5°C, to avoid visually significant refractive error due to calibration error. Further experimental investigations are required to determine the influence of different initial corneal temperatures on the refractive outcome.
18