Rapid Method for Measuring Rotenone in Water at Piscicidal Concentrations

Verdel K. Dawson, Paul D. Harman, David P. Schultz, John Allen

Transactions of the American Fisheries Society · 1983 · 13 citations · 2 references

Concepts

Abstract

Abstract A high-performance liquid chromatography (HPLC) procedure that is rapid, specific, and sensitive (limit of detection <0.005 mg/liter) was developed for monitoring application and degradation rates of rotenone. For analysis, a water sample is buffered to pH 5 and injected through a Sep Pak(R) C18 disposable cartridge. The cartridge adsorbs and retains the rotenone which then can be eluted quantitatively from the cartridge with a small volume of methanol. This step effectively concentrates the sample and provides sample cleanup. The methanol extract is analyzed directly by HPLC on an MCH 10 reverse-phase column; methanol: Water (75:25, volume : Volume) is the mobile phase and flow rate is 1.5 ml/minute. The rotenone is detected by ultraviolet spectrophotometry at a wavelength of 295 nm. Received November 29, 1982 Accepted May 24, 1983

References

2