Diagnostic Test Validation

Leonardo Elías Ordóñez Ordóñez, Eveling Rojas-Roncancio, Víctor Hernández-Alarcón, Rafael Jaramillo-Safón, José A. Prieto-Rivera, José E. Guzmán-Durán, José G. Lora-Falquez, Marcela Escobar, Esther Sofía Angulo‐Martínez

Otology & Neurotology · 2009 · 13 citations · 11 references

Concepts

Abstract

OBJECTIVE: To determine the diagnostic value of the cochlear hydrops analysis masking procedure (CHAMP) in patients with definite Ménière's disease. STUDY DESIGN: Prospective validation study for a diagnostic test, study phase 2. SETTING: Tertiary referral centers. PATIENTS: Subjects with definite Ménière's disease (Group 1), differential diagnosis (Group 2: another audiovestibular diseases or neurologic disorders), and normal hearing (Group 3) were included. Sample sizes were calculated. INTERVENTIONS: Study test (CHAMP) was compared with the current clinical criterion standard described by the guidelines of the American Academy of Otolaryngology-Head and Neck Surgery. MAIN OUTCOME MEASURE: Sensitivity, specificity, and other indicators of diagnostic validation. RESULTS: One hundred ten cases completed the follow-up, and their results are presented. Sensitivity at 31.3% and specificity at 100% were found in subjects with definite Ménière's disease, features that are more helpful in confirming the diagnosis than in rejecting it. Group 1 showed significantly shorter latency delays than Groups 2 and 3 (p < 0.001). CONCLUSION: If definite Ménière's disease is suspected, an abnormal result confirms the diagnosis; however, a normal result does not rule out the Ménière's disease diagnosis.

References

11