An Evaluation of 13 Commercial Kits for the Measurement of Testosterone in Serum and Plasma

M.J. Wheeler, M Shaikh, R. D. Jennings

Annals of Clinical Biochemistry International Journal of Laboratory Medicine · 1986 · 12 citations · 7 references

Concepts

Abstract

An evaluation has been carried out of 13 commercial testosterone kits. The within- and between-assay imprecision was found to be unsatisfactory for some kits. Sensitivity and linearity were acceptable for all the kits but several showed concentration-dependent biases to the ALTM or GCMS values. The occurrence of anomalous results and changing performance characteristics of some of the direct kits is also presented.

References

7