Publication | Open Access
Evaluation of thermal injury to liver, pancreas and kidney during irreversible electroporation in an <i>in vivo</i> experimental model
110
Citations
29
References
2014
Year
IRE can generate thermal energy, and even thermal injury, based on tissue type, probe exposure lengths, pulse lengths and proximity to metal. Awareness of probe placement regarding proximity to critical structures as well as probe exposure length and pulse length are necessary to ensure safety and prevent thermal injury. A probe exposure of 2·5 cm or less for liver IRE, and 1·5 cm or less for pancreas, with maximum pulse length of 90 µs will result in safe and non-thermal energy delivery with spacing of 1·5-2·3 cm between probe pairs.
| Year | Citations | |
|---|---|---|