Audiology and Neurotology · 2012 · 54 citations · 13 references
Health ProfessionsBiomedical EngineeringOtolaryngologyAuditory AnatomyCochlear ImplantsOtologyBasilar Membrane AnatomyCochlear AnatomyCochlear ImplantationCochlear MechanicsImplant TechnologySurgical ImplantationCochlear Electrode ArraysCochlear PhysiologyElectrode Array InsertionInsertion Force AnalysisInsertion Force MeasurementInsertionsInsertions DepthIntracochlear Trauma
Cochlear implant array insertion forces are potentially related to cochlear trauma. We compared these forces between a standard (Digisonic SP; Neurelec, Vallauris, France) and an array prototype (Neurelec) with a smaller diameter. The arrays were inserted by a mechatronic tool in 23 dissected human cochlea specimens exposing the basilar membrane. The array progression under the basilar membrane was filmed together with dynamic force measurements. Insertion force profiles and depth of insertion were compared. The recordings showed lower insertion forces beyond 270° of insertion and deeper insertions with the thin prototype array. This will potentially allow larger cochlear coverage with less trauma.
13
Conservation of low-frequency hearing in cochlear implantation
Jan Kiefer, Wolfgang Gstœttner, Stephen Marcel Pok et al. · Acta Oto-Laryngologica · 2004 · 303 citations
Impact of electrode insertion depth on intracochlear trauma
Oliver F. Adunka, Jan Kiefer · Otolaryngology · 2006 · 190 citations