Real-time proton beam range monitoring by means of prompt-gamma detection with a collimated camera

F. Roellinghoff, A. Benilov, D. Dauvergne, G. Dedes, N. Freud, Guillaume Janssens, J. Krimmer, Jean Michel Létang, Marco Pinto, D. Prieels,

Physics in Medicine and Biology · 2014 · 60 citations · 17 references

Concepts

Abstract

Prompt-gamma profile was measured at WPE-Essen using 160 MeV protons impinging a movable PMMA target. A single collimated detector was used with time-of-flight (TOF) to reduce the background due to neutrons. The target entrance rise and the Bragg peak falloff retrieval precision was determined as a function of incident proton number by a fitting procedure using independent data sets. Assuming improved sensitivity of this camera design by using a greater number of detectors, retrieval precisions of 1 to 2 mm (rms) are expected for a clinical pencil beam. TOF improves the contrast-to-noise ratio and the performance of the method significantly.

References

17