Efficiency Measurements and Installation of a New Grating for the OSIRIS Spectrograph at Keck Observatory

Etsuko Mieda, S. Wright, James Larkin, James R. Graham, Sean M. Adkins, James E. Lyke, Randy Campbell, Jérôme Maîre, Tuan Do, Jacob Gordon

Publications of the Astronomical Society of the Pacific · 2014 · 34 citations · 19 references

DOIFull text

Open access

Concepts

Abstract

OSIRIS is a near-infrared integral field spectrograph operating behind the adaptive optics system at W. M. Keck Observatory. While OSIRIS has been a scientifically productive instrument to date, its sensitivity has been limited by a grating efficiency that is less than half of what was expected. The spatially averaged efficiency of the old grating, weighted by error, is measured to be 39.5 +/- 0.8 % at {\lambda} = 1.310 {\mu}m, with large field dependent variation of 11.7 % due to efficiency variation across the grating surface. Working with a new vendor, we developed a more efficient and uniform grating with a weighted average efficiency at {\lambda} = 1.310 {\mu}m of 78.0 +/- 1.6 %, with field variation of only 2.2 %. This is close to double the average efficiency and five times less variation across the field. The new grating was installed in December 2012, and on- sky OSIRIS throughput shows an average factor of 1.83 improvement in sensitivity between 1 and 2.4 microns. We present the development history, testing, and implementation of this new near-infrared grating for OSIRIS and report the comparison with the predecessors. The higher sensitivities are already having a large impact on scientific studies with OSIRIS.

References