Journal of The Electrochemical Society · 1999 · 28 citations · 0 references
PharmacyChemical EngineeringElectroanalytical MethodsElectrochemical EngineeringElectrochemistryElectrochemical DetectionElectrochemical SensorsIon-selective ElectrodesFlow SensitivityHigh-sensitivity AssaysPolymer ElectrodesPolypyrroleSample InjectionThin-film Composite StructureUrea BiosensorsUrea-polymer Complexes
A highly sensitive and rapid flow injection system for urea analysis was constructed with a composite film of electropolymerized inactive polypyrrole (PPy) and a polyion complex incorporating urease. This system shows a sensitivity of and a lifetime of more than 80 assays. The origin of the high sensitivity of this system is attributed to an additional potential response of inactive PPy to ammonia or ammonium ion superimposed on the response to pH change. By injecting a concentrated buffer solution immediately after the sample injection, this system is capable of assaying more than 15 samples per hour. © 1999 The Electrochemical Society. All rights reserved.