Lab on a Chip · 2007 · 77 citations · 35 references
EngineeringBioengineeringBiotechnologyBiosensorsElectroanalytical MethodsSensorsDna BiosensorsElectrochemical SensorsMicrofluidic BiosensorsMicrofluidic DevicesMicrofluidicsNucleic Acid BiosensorsConductanceDna SensorsImmobilizationInterdigitated Electrode SensorsLabel-free DetectionLabel-free Nucleic Acid DetectionPathogen DnaPathogen Dna Detection
We investigate the analytical performance of an interdigitated electrode sensor for the label-free detection of DNA, by monitoring the complex impedance of 5 microm wide interdigitated Pt microelectrodes on a glass substrate. We detect the hybridization of unlabeled 38-mer target ssDNA with a complementary probe that is bound on the glass in between the electrodes by a disuccinimidyl terephtalate and aminosilane immobilization procedure. The sensor is mounted in a microfluidic flow cell, in which hybridization is monitored and in situ compared with a reference. After hybridization, the cell is perfused with deionised water and the dependence of the measured conductance due to the immobilized target DNA layer, to target DNA concentrations down to 1 nM is demonstrated. Subsequently, we apply our sensor to the detection of pathogen DNA from Salmonella choleraesuis in dairy food.
35
Array-Based Electrical Detection of DNA with Nanoparticle Probes
So‐Jung Park, T. Andrew Taton, Chad A. Mirkin · Science · 2002 · 1.5K citations
Mutation detection by electrocatalysis at DNA-modified electrodes
Elizabeth M. Boon, Donato M. Ceres, T. Gregory Drummond et al. · Nature Biotechnology · 2000 · 636 citations