International Journal of Environmental & Analytical Chemistry · 1992 · 20 citations · 3 references
EngineeringChemistryAir PollutionAnalytical ChemistryInstrumentationAir Pollution InstrumentationAir Pollution MeasurementAir SamplingAnalytical MeasurementsGas SensorsVolatile Organic CompoundsCarrier GasesDynamic CalibrationGas DetectionGas Detection InstrumentsGas Detection TechnologyGas Phase Pollutant SamplingGravimetric MeasurementOccupational Hygiene Engineering
Abstract This paper describes the design, performance and operation of a home-made capillary dosage unit with in situ weight sensor. This continuous injection technique is developed to produce dynamic gas phase calibration mixtures of VOC's in a diluent gas at occupational hygiene levels. A new advance in the technique allows the calculation of the amounts of pollutants emitted into the carrier gas, directly from the gravimetric measurements. This is achieved by a built-in weight sensor which is monitored without interfering with the injection process. The survey of the capillary dosage unit(s) is managed by a microcomputer which automatically lists the amounts of pollutants introduced into the diluent gas. The capillary dosage technique thus becomes related directly to a primary standard. The homogeneity and stability of the test gas generated is shown to be better than 1% relative standard deviation over periods of up to 24 hours. The VOC concentrations in the manifold are known with a total uncertainty of between 2 and 3%SS.
3
Primary Standards for Trace Gas Analysis.
Andrew E. O'Keeffe, Gordon C. Ortman · Analytical Chemistry · 1966 · 295 citations