In Vitro Simulation

Allen Carl, Harpal S. Khanuja, Barton L. Sachs, Charles A. Gatto, John vomLehn, Kirby G. Vosburgh, John F. Schenck, William E. Lorensen, David G. Disler

Spine · 1997 · 80 citations · 20 references

Concepts

Abstract

STUDY DESIGN: Frameless stereotaxy with doppler ultrasound and three dimensional computer model registration is assessed in vitro for pedicle screw placement. OBJECTIVE: To identify feasibility of pedicle screw navigation and placement using this technology. SUMMARY OF BACKGROUND DATA: Inaccurate pedicle screw placement can lead to neurovascular injury or suboptimal fixation. Present techniques in pedicle screw placement involve only confirmation of hole orientation. METHOD: Forty-four pedicle screws were placed in lumbosacral models and cadaver specimens. Accuracy was assessed with a computed tomography scan and vertebral cross sectioning. RESULTS: All screws were intrapedicular. Accuracy of anterior cortical fixation was 1.5 mm, with a range of 2.5 mm. CONCLUSION: In vitro frameless stereotaxy is accurate for pedicle screw placement. This technology adds a component of navigation to pedicle screw placement.

References

20