Publication | Closed Access
Full-field optical coherence tomography with thermal light
76
Citations
10
References
2002
Year
Thermal LightEngineeringMicroscopyPiezoelectric TranslatorBiomedical EngineeringOptical PropertiesLight MicroscopyRadiologyHealth SciencesMedical ImagingOphthalmologyFree SpaceBiophotonicsOptical ImagingReference MirrorBiomedical ImagingOptical Coherence TomographyQuantitative Phase ImagingTomography
A simple optical coherence tomography system has been developed based on a white-light Linnik interferometric microscope with its reference mirror mounted on a piezoelectric translator. The geometrical extension of the optics allows efficient illumination of this device with a low-power (3-W) light bulb, yielding full-field interferometric images at 50 Hz with a fast CCD camera. Owing to the very broad spectral width of the light source and of the camera response, we achieved axial resolutions equal to 1.1 microm in free space and 0.7 microm through a standard microscope cover plate. Tomographic images of an epithelial cell smear and of an hematological sample are shown.
| Year | Citations | |
|---|---|---|