Publication | Open Access
Comparison of the glidescope, CMAC, storz DCI with the Macintosh laryngoscope during simulated difficult laryngoscopy: a manikin study
33
Citations
22
References
2012
Year
Unsurprisingly, when used in a simulated difficult laryngoscopy, all the videolaryngoscopes resulted in a better view of the glottis than the Macintosh blade. However, interestingly the CMAC was found to provide a better laryngoscopic view that the Storz DCI Videolaryngoscope. Additionally, use of either the Glidescope or Storz DCI Videolaryngoscope resulted in a prolonged time to successful intubation compared with use of the CMAC or Macintosh blade. The use of the CMAC during manikin simulated difficult laryngoscopy combined the efficacy of attainment of laryngoscopic view with the expediency of successful intubation. Use of the Macintosh blade combined expedience with success, despite a limited laryngoscopic view. The limitations of a manikin model of difficult laryngoscopy limits the conclusions for extrapolation into clinical practice.
| Year | Citations | |
|---|---|---|
Page 1
Page 1