2013 · 22 citations · 7 references
EngineeringBiomedical EngineeringComputer EngineeringBiomedical Signal ProcessingLow-power DesignSensorsData CompressionDigital Signal ProcessingElectrocardiographyLossless Data CompressionLossy Data CompressionLow-power Digital Circuit DesignCompression RatioEcg Signal CompressionElectrocardiogram CompressionEstimation ErrorFixed-length CodingPower ConsumptionUltra-low-power CmosWearable Sensors
This paper presents a low power ECG recording Sys-tem-on-Chip (SoC) with on-chip low complexity lossless ECG compression for data reduction in wireless/ambulatory ECG sensor devices. The proposed algorithm uses a linear slope predictor to estimate the ECG samples, and uses a novel low complexity dynamic coding-packaging scheme to frame the resulting estimation error into fixed-length 16-bit format. The proposed technique achieves an average compression ratio of 2.25× on MIT/BIH ECG database. Implemented in 0.35 μm process, the compressor uses 0.565 K gates/channel occupying 0.4 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> for 4-channel, and consumes 535 nW/channel at 2.4V for ECG sampled at 512 Hz. Small size and ultra-low power consumption makes the proposed technique suitable for wearable ECG sensor application.