Multichannel millimeter wave interferometer for W7-AS
20
Citations
4
References
1997
Year
EngineeringMultichannel MillimeterPhysicsGunn OscillatorCalibrationLocal OscillatorNatural SciencesMillimeter WaveInterferometryTen ChannelRadio FrequencyMicrowave MeasurementMicrowave DiagnosticsInstrumentationMillimeter Wave TechnologyRadio TelescopeAstrophysics
A newly designed ten channel microwave interferometer for line integrated electron density measurements at the experiment W7-AS stellarator is described. The probing signals in the frequency range of 160–162 GHz are generated by Gunn oscillators with subsequent frequency doublers. Double conversion heterodyne receivers with 2 MHz bandwidth are used for detection. All oscillators involved are phase locked. Use of each Gunn oscillator is twofold: as a source for the probing signal and as a local oscillator for heterodyne detection. The dynamic range of the millimeter wave part is about 70 dB. The instrument operates up to a peak density of 7×1019 m−3 with a temporal resolution of 5 μs. This article gives a detailed technical description of the system and demonstrates its usefulness by means of an experimental example.