Journal of Medical Engineering & Technology · 2013 · 13 citations · 20 references
MedicineOrthopedic SurgerySurgeryHemodynamic ManagementSurgical InstrumentationVascular SurgeryBlood Pressure ControlBlood Pressure Signal AnalysisCardiovascular MonitoringHemodynamicsArterial Blood Pressure MeasurementBlood PressureContinuous Arterial Pressure MonitoringContinuous Blood Pressure MonitoringMean Arterial Pressure EstimationNoninvasive Blood Pressure MeasurementSystolic Blood PressureSystolic Blood Pressure Estimation
The Tensys T-line uses tonometry to reproduce the arterial blood pressure tracing non-invasively. The purpose of this study was to assess the agreement between estimates of the T-line and an intra-arterial catheter (for both mean arterial pressure [MAP] and pulse pressure variation [PPV]) in the setting of spine surgery. Continuous blood pressure data were collected for 7507 minutes from 25 patients. Five increasingly aggressive T-line filters were applied. The mean bias for mean, diastolic and systolic blood pressure ranged from 3.4-6.4, 3.1-7.1 and 0.1-0.8 mmHg and 6.5-11.8% for PPV. Ninety-five per cent confidence intervals for mean, diastolic and systolic blood pressure ranged from 24-28, 23.1-24.7 and 33.4-35.6 mmHg for 14-21% for PPV. The limits of agreement preclude the use of the T-line for reliable estimation of MAP or PPV in spine surgery.
20
A H BECKETT, M ROWLAND, P TURNER · The Lancet · 1965 · 16.9K citations
Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®
Jerson R. Martina, Berend E. Westerhof, Jeroen van Goudoever et al. · Anesthesiology · 2012 · 291 citations
Medicine, Cardiology, Surgery +16