Determination of lithium in microlitre amounts of human body fluids at therapeutic and normal levels by stable isotope dilution and field desorption mass spectrometry

Wolf D. Lehmann, U. Bahr, H.‐R. Schulten

Journal of Mass Spectrometry · 1978 · 30 citations · 13 references

Concepts

Abstract

The use of field desorption mass spectrometry for the determination of lithium in body fluids at therapeutic levels--ppm region--as well as at the normal level--ppb region--has been developed. The use of a stable isotope enriched internal standard, together with the outstanding sensitivity of field desorption for alkali cations and the high specificity of mass spectrometry, allows a quantitative determination of lithium in microlitre amounts of body fluids, such as plasma, saliva and urine. The assay allows a determination of lithium even at ultratrace concentrations where routine spectroscopic procedures cannot be applied. Analysis of plasma required a simple protein precipitation, whereas saliva and urine could be analysed without treatment. The precision of the data obtained ranged from 2--10%. The time consumption for one analysis in routine work is about 20--30 min.

References

13