Ghosting comparison for large-area selenium detectors suitable for mammography and general radiography
21
Citations
8
References
2004
Year
Dynamic ImagingEngineeringRadiation DetectionMedical ImagingMeasurement24X30cm DetectorDetector PhysicInstrumentationDetection TechnologyRadiation ImagingGeneral RadiographyNew 43X43cmNuclear MedicineLarge-area Selenium DetectorsRadiologyHealth Sciences
In this paper, two new selenium large area detectors are introduced. The first detector is a 24x30cm detector suitable for full-field digital mammography applications. The second detector is a new 43x43cm detector for general radiographic applications. Both detectors are capable of static and dynamic imaging, and are quantum noise limited over the exposure ranges intended for their typical use. For static imaging applications, ghost and lag were compared on both detectors, and no measurable artefacts were reported. On dynamic imaging sequences, lag was shown to be significant on both detectors, and a method for reducing the artifact due to lag was presented.