Compliance as a factor effecting the patency of a copolyurethane vascular graft
96
Citations
5
References
1978
Year
Tissue EngineeringEngineeringSurgeryBiomedical EngineeringVascular SurgeryGraft SurvivalVascularized Bone GraftAtherosclerosisVascular AdaptationNatural ArteryVascular BiologyCompliant GraftsCardiovascular DiseasePhysiologyCopolyurethane Vascular GraftVascular AccessSoft Tissue ReconstructionMedicinePlastic SurgeryLongest Implant TimeAnesthesiology
A new solid-wall vascular graft which has a compliance approximating that of the natural artery has been prepared from a copolyether-urethane material. Six of nine of these compliant grafts implanted in dogs were patent on removal, the longest implant time being 77 days for a 4-mm I.D. femoral artery graft. This is in contrast to noncompliant grafts of the same copolyurethane in which failure usually occurred within 48 hr.