Review of Scientific Instruments · 1996 · 71 citations · 10 references
Natural SciencesChemistryAnalytical ChemistryInstrumentationMass Spectrometric AnalysisAccurate Mass MeasurementMass SpectrometryFragmentation PathwaysHigh-resolution Mass AnalysisTandem Mass Spectrometry HardwareTime-of-flight Mass Spectrometry
Modulation of an ion beam is crucial to several applications in time-of-flight (TOF) mass spectrometry, especially for tandem TOF instruments which require selection of a particular precursor m/z (mass to charge) value prior to fragmentation. Here we present a detailed description of an ‘‘interleaved comb’’ ion deflection gate device with suitable electronics which offers a performance advantage over the more commonly used deflection plate devices. We demonstrate unit mass resolution for selection to m/z 167 in the tandem TOF instrument constructed in our laboratory. Calculations suggest that the real time of unit resolution for our experimental arrangement is greater than 300.
10
<i>Basic Processes of Gaseous Electronics</i>
Leonard B. Loeb, David J. Rose · Physics Today · 1956 · 621 citations
Absolute Values of the Electron Mobility in Hydrogen
Norris E. Bradbury, Russell A. Nielsen · Physical Review · 1936 · 339 citations