Analytical Methods · 2010 · 30 citations · 41 references
Natural SciencesChemistryAir PollutionAnalytical ChemistryMass Spectrometric AnalysisEnvironmental Analytical ChemistryInstrumental AnalysisIon ChromatographyMass SpectrometryParticulate MatterRadiochemical AnalysisActinide RadiochemistryAir FiltersAnalytical Detection LimitsFiltersHigh-efficiency Particulate Air (Hepa) FiltersMass Spectrometric Isotope AnalysisSpectral Interferences
A novel, rapid and completely automated method for the determination of long-lived actinides in air particulate samples was developed using ion chromatography separation prior to ICP-MS analyses. U, Pu and Am were pre-concentrated and separated from radioactive and stable interferences through a sequential arrangement of three columns (TEVA, U/TEVA, DGA) arranged in order to allow selective retention of the radioisotope on the designated resin. Detection and quantification of the various radioisotopes were performed on-line using inductively coupled plasma mass spectrometry (ICP-MS). Detection limits of 0.0006 (238U), 0.0063 (239Pu), 0.0041 (240Pu), 0.062 (241Am) μBq/m3 were obtained for air filters (based on 3000 m3 of air filtered). The automated procedure has been successfully tested on SRMs and spiked samples and showed high recovery yields (>90%). A total analysis time of 18 min is required for the separation, detection, and column rinsing/regeneration for subsequent analysis.
41