High-power ultrasonic dissections at level 5 and to a lesser extent level 4 result in considerable heat production that causes proximity collateral damage to adjacent tissues when the continuous activation time exceeds 10 seconds. Ultrasonic dissections near important structures should be conducted at level 3. At power levels of 4 and 5, the ultrasonic energy bursts to the tissue should not exceed 5 seconds at any one time.