Ability of Two Commercially Available Assays (Abbott RealTi <i>m</i> e HIV-1 and Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Version 2.0) To Quantify Low HIV-1 RNA Levels (&lt;1,000 Copies/Milliliter): Comparison with Clinical Samples and NIBSC Working Reagent for Nucleic Acid Testing Assays

Alessandra Amendola, Patrizia Marsella, Maria Bloisi, Federica Forbici, Claudio Angeletti, Maria Rosaria Capobianchi

Journal of Clinical Microbiology · 2014 · 30 citations · 38 references

DOIFull text

Open access

Concepts

Abstract

Concordance between molecular assays may be suboptimal at low HIV-1 viremia levels (<1,000 copies/ml); therefore, it may be difficult to define and compare virologic endpoints for successful and failed therapy. We compared two commercial assays (the Abbott RealTime HIV-1 and the Roche Cobas AmpliPrep/TaqMan HIV-1 version 2.0) for their ability to detect and quantify low viral loads. A comparison was performed using 167 residual clinical samples (with values ranging from "not detected" to 1,000 copies/ml, as measured by the Abbott assay) and the National Institute and Biological Standards and Control (NIBSC) HIV-1 RNA working reagent 1 for nucleic acid amplification techniques (NAT) assays (serially diluted to a range from 1 to 1,000 copies/ml). Quantitative results were compared using Lin's concordance correlation coefficient and a Bland-Altman plot. Concordance with the qualitative results was measured by Cohen's kappa statistic. With clinical samples, the degree of interassay concordance of the qualitative results at a 40-copies/ml HIV-1 RNA threshold was substantial (κ = 0.762); the correlation among the quantified samples was suboptimal (concordance correlation coefficient, 0.728; P < 0.0001); the mean difference of the values between the Roche and Abbott assays was 0.193 log10 copies/ml. Using the HIV-1 RNA working reagent 1 for NAT assays, the results provided by the Roche assay were, on average, 3 times higher than expected, while the Abbott assay showed high accuracy. The Roche assay was highly sensitive, being able to detect a level as low as 3.5 copies/ml HIV-1 RNA with 95% probability. The performance characteristics of each molecular assay should be taken into account when HIV-1 RNA threshold values for "virologic suppression," "virologic failure," "persistent low viral loads," etc., are defined and indicated in the support of clinical decisions.

References

38