Journal of Analytical Atomic Spectrometry · 1988 · 18 citations · 0 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElemental AnalysisEnvironmental ChemistryAnalytical MeasurementsAtomic Emission SpectroscopyEnvironmental AnalysisEnvironmental Analytical ChemistryHeavy Metal AnalysisInstrumental AnalysisAnalytical Detection LimitsAtomic Emission SpectrometryDetectionInductively Coupled Plasma Atomic Emission SpectroscopyInductively Coupled Plasma SpectrometryLong-term StabilityOptimization Of Operating ConditionsSample Introduction Systems
A thermospray nebuliser, used for sample introduction into an ICP, was equipped with a heated spray chamber in order to enhance the detection power of the sample introduction system. The carrier gas flow, the aerosol (desolvation) temperature and the condenser temperature were optimised. At the optimised operating conditions about 10% of the nebulised analyte reaches the ICP. The detection limits obtained are comparable to those obtained using ultrasonic nebulisers. Both short- and long-term stability and matrix effects are discussed. Finally, data obtained for the determination of chromium, cadmium and silver in the standard samples EPA WPQCS2 and NBS SRM1643A show the analytical potential of the proposed system.