Publication | Closed Access
Anodal Right Ventricular Capture During Left Ventricular Stimulation in CRT‐Implantable Cardioverter Defibrillators
37
Citations
9
References
2005
Year
RV anodal capture during LV pacing was found in 11/11 patients at some output with true bipolar RV leads versus 0/15 patients with RV leads with an integrated bipolar design. Anodal RV capture threshold was more affected by changes in pw duration than LV capture threshold. In CRT-ICD systems, RV leads with a true bipolar design with the proximal ring also used as the anode for LV pacing are associated with a high incidence of anodal RV capture during LV pacing. This may affect the clinical response to alternative resynchronization methods using single LV stimulation or interventricular delay programming.
| Year | Citations | |
|---|---|---|
Page 1
Page 1