Chinese Physics C · 2010 · 12 citations · 7 references
Natural SciencesPhysicsAnalytical ChemistryInstrumentationNuclear PhysicsAccurate Mass MeasurementCharged Particle DetectionDetector SystemsDetector TechnologyMass MeasurementMass SpectrometryHigh-resolution Mass AnalysisSecondary Electron DetectorsSpectrometer DesignSpectrometer ResolutionTime Resolution
The Isochronous Mass Spectrometry is a high accurate mass spectrometer. A secondary electrons time detector has been developed and used for mass measurements. Secondary electrons from a thin carbon foil are accelerated by an electric field and deflected 180° by a magnetic field onto a micro-channel plate. The time detector has been tested with alpha particles and a time resolution of 197 ps (FWHM) was obtained in the laboratory. A mass resolution around 8×10−6 for Δm/m was achieved by using this time detector in a pilot mass measurement experiment.