Flow Rate-Pressure Drop Relation in Coronary Angioplasty: Catheter Obstruction Effect
75
Citations
16
References
1996
Year
HypertensionHeart FailureEndovascular TechniqueSurgeryCoronary Artery DiseaseBlood FlowVascular SurgeryBlood Flow MeasurementCardiologyPercutaneous Coronary InterventionFlow BlockageCatheter PresentCardiovascular DiseaseCatheter Obstruction EffectVascular AccessMedicineAngioplasty CatheterEmergency MedicineAnesthesiology
Quantitative methods to measure the hemodynamic consequences of various endovascular interventions including balloon angioplasty are limited. Catheters measuring translesional pressure drops during balloon angioplasty procedures can cause flow blockage and thus inaccurate estimates of pre- and post-intervention flow rates. The purpose of this investigation was to examine the influence of the presence and size of an angioplasty catheter on measured mean pressure gradients across human coronary artery stenoses. Analytical flow modeling and in vitro experimental evidence, coupled with angiographic data on the dimensions and shape of stenotic vessel segments before and after angioplasty, indicated significant flow blockage effects with the catheter present.