Journal of Biomedical Optics · 2003 · 32 citations · 27 references
EngineeringBioengineeringBiomedical EngineeringBiomedical OpticsOptoelectronicsSensorsHemoglobin PhysiologyInfrared InstrumentationInfrared SpectroscopyNear-infrared SpectroscopyOptical BiosensingOptical BiosensorsArterial Oxygen SaturationOptical Sensor CalibrationOxygen SaturationOxygen Saturation MeasurementOxygen Saturation MonitoringPulse OximetersPulse OximetryWavelength Selection
Pulse oximeters are widely used for noninvasive monitoring of oxygen saturation in arterial blood hemoglobin. We present a transmittance pulse oximetry system based on near-infrared (NIR) laser diodes (750 and 850 nm) for monitoring oxygen saturation of arterial blood hemoglobin. The pulse oximetry system is made up of the optical sensor, sensor electronics, and processing block. Also, we show experimental results obtained during the development of the whole NIR transmittance pulse oximetry system along with modifications in the sensor configuration, signal processing algorithm, and calibration procedure. Issues concerning wavelength selection and its implications for the improvement of the transmittance pulse oximetry technique are discussed. The results obtained demonstrate the proposed system's usefulness in monitoring a wide range of oxygen saturation levels.
27
A H BECKETT, M ROWLAND, P TURNER · The Lancet · 1965 · 16.9K citations
Optical Constants of Water in the 200-nm to 200-μm Wavelength Region
George M. Hale, Marvin R. Querry · Applied Optics · 1973 · 4.4K citations