Pericardial Effusion and Tamponade in Infants With Central Catheters
205
Citations
35
References
2002
Year
The pericardial fluid found in CVC-associated PCE is consistent with the infusate. We speculate that there are several mechanisms, ranging from frank perforation that seals spontaneously to CVC tip adhesion to the myocardium with diffusion into the pericardial space. Routine radiography should be performed, and the CVC tip should be readily identifiable. The CVC tip should remain outside the cardiac silhouette but still within the vena cavae (approximately 1 cm outside the cardiac silhouette in premature infants and 2 cm in term infants). A change in cardiothoracic ratio may be diagnostic of a PCE, and pericardiocentesis is associated with significantly reduced mortality. Increased awareness of this complication may decrease the mortality associated with CVC-related PCE.